Sulfuric acid solution

c(H₂SO₄) = 0.005 mol/l (0.01 N), reag. USP, ready-to-use volumetric solution for titration, Titriplex®

Synonym(e):

Schwefelsäure -Lösung

Anmeldenzur Ansicht organisationsspezifischer und vertraglich vereinbarter Preise


About This Item

Lineare Formel:
H2SO4
CAS-Nummer:
Molekulargewicht:
98.08
Beilstein:
2037554
MDL-Nummer:
UNSPSC-Code:
12352322
NACRES:
NA.21

product name

Schwefelsäure, c(H2SO4) = 0.005 mol/l (0.01 N), Titripur®, reag. USP

Agentur

reag. USP

Qualitätsniveau

Produktlinie

Titripur®

Form

liquid

Qualität

Analyzed in our ISO 17025 accredited QC lab

Eignung der Reaktion

reaction type: Acid-base reactions

Konzentration

0.005 M

Methode(n)

titration: suitable

pH-Wert

2 (20 °C in H2O)

Dichte

0.999 g/cm3 at 20 °C

Lagertemp.

15-25°C

SMILES String

OS(O)(=O)=O

InChI

1S/H2O4S/c1-5(2,3)4/h(H2,1,2,3,4)

InChIKey

QAOWNCQODCNURD-UHFFFAOYSA-N

Suchen Sie nach ähnlichen Produkten? Aufrufen Leitfaden zum Produktvergleich

Leistungsmerkmale und Vorteile

Accreditation:
This volumetric solution is analyzed by our calibration laboratory D-K-15185-01-00 which is accredited according to DIN EN ISO/IEC 17025 for analysis of amount-of-substance concentrations in volumetric solutions by DAkkS (Deutsche Akkreditierungsstelle - German National Accreditation Body). The accreditation certificate can be found at www.sigmaaldrich.com/ISO17025.

Hinweis zur Analyse

Form liquid
Amount-of-substance concentration 0.004975 - 0.005025 mol/L
Measurement uncertainty ± 0.000015 mol/L
Traceability NIST SRM

The concentration is determined by volumetric titration and refers to 20°C.
The amount-of-substance concentration of this volumetric solution is traceable to a primary standard reference material (SRM) from the National Institute of Standards and Technology, Gaithersburg, USA (NIST SRM 723 Tris(hydroxymethyl)aminomethane) by means of volumetric standard Tris(hydroxymethyl)aminomethane (article number 1.02408), certified reference material according to ISO 17034, analyzed by our accredited calibration laboratory of Merck KGaA, Darmstadt, Germany according to DIN EN ISO/IEC 17025. The uncertainty is expressed as expanded measurement uncertainty with a coverage factor k=2 covering a confidence level of 95%.

Note: The titer is a correction factor to correct for variations of the volumetric solution, the titration equipment, the temperature and other laboratory conditions. For correct titration results it is recommended to determine a titer with the laboratory specific equipment and under laboratory specific conditions directly after opening a new bottle and at regular time intervals.

Rechtliche Hinweise

Titripur is a registered trademark of Merck KGaA, Darmstadt, Germany

Lagerklassenschlüssel

12 - Non Combustible Liquids

WGK

nwg

Flammpunkt (°F)

Not applicable

Flammpunkt (°C)

Not applicable


Analysenzertifikate (COA)

Suchen Sie nach Analysenzertifikate (COA), indem Sie die Lot-/Chargennummer des Produkts eingeben. Lot- und Chargennummern sind auf dem Produktetikett hinter den Wörtern ‘Lot’ oder ‘Batch’ (Lot oder Charge) zu finden.

Besitzen Sie dieses Produkt bereits?

In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.

Die Dokumentenbibliothek aufrufen

Kunden haben sich ebenfalls angesehen

Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..

Setzen Sie sich mit dem technischen Dienst in Verbindung.