G005594
HPLC Analysis of Maté Leaves on Ascentis® Express C18
application for HPLC
About This Item
tecniche
HPLC: suitable
Parametri analitici
sample/matrix: UAE (Ultrasonic Extraction) maté extract (dried and minced leaves (7.5 g) from maté were extracted with 225 mL of methanol in an ultrasonic bath, thermostated at 75 ± 0.5 °C (potency: 90 W; frequency: 40 kHz))
column: Ascentis Express C18, 150 x 4.6 mm I.D., 2.7 μm particles (53829-U)
mobile phase: [A] water, pH 3 with formic acid; [B] acetonitrile, pH 3 with formic acid
gradient: 0 to 5% B in 20 min; 5 to 40% B in 30 min
flow rate: 1 mL/min
column temp.: 25 °C
detector: PDA: range 215- 420 nm (8 μL detector cell volume, cell temp. 40 °C, sampling rate 12.5 Hz, time constant 0.025 s)
injection: 10 μL
Compatibilità
application for HPLC
applicazioni
vitamins, nutraceuticals, and natural products
Note legali
Colonna analitica
Standard
Choose from one of the most recent versions:
Certificati d'analisi (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Servizio Clienti
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..