G005524
US EPA Method 524.2: GC Analysis of Volatiles on SPB-624 after Purge & Trap using "K" Trap
suitable for GC
About This Item
Prodotti consigliati
tecniche
gas chromatography (GC): suitable
Parametri analitici
sample/matrix: each analyte at 10 ppb in 25 mL water
purge trap: VOCARB 3000 "K" (24940-U)
dry purge: 2 min
desorption preheat: 205 °C
desorption process: 40 mL/min at 210 °C for 2 min
bake: 260 °C for 10 min
transfer line/valve temp.: 110 °C
sample preparation: purge is 40 mL/min at 25 °C for 11 min
column: SPB-624, 30 m x 0.25 mm I.D., 1.4 μm (24255)
oven: 40 °C (2 min), 7 °C/min to 135 °C, 30 °C/min to 230 °C (3 min)
carrier gas: helium, 1.2 mL/min
injection: 30:1 split
liner: 0.75 mm I.D. SPME