GC Analysis of Terpenes in Hops (Cascade Variety) on Equity-1 after Headspace SPME using 50/30 μm DVB/CAR/PDMS Fiber

GC Analysis of Terpenes in Hops (Cascade Variety) on Equity-1 after Headspace SPME using 50/30 μm DVB/CAR/PDMS Fiber suitable for GC, application for SPME

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

Equity-1

L × I.D. 60 m × 0.25 mm, df 0.25 μm

Fibre SPME

Réf. du produit
Description
Tarif

Fibres SPME Divinylbenzene/Carboxen/Polydimethylsiloxane (DVB/CAR/PDMS)

needle size 23 ga, StableFlex, for use with autosampler

À utiliser avec

Réf. du produit
Description
Tarif

SPME/Direct

pkg of 1 ea

Flacons Headspace, haut à vis, fond arondi (flacon seul)

volume 10 mL, clear glass vial, thread for 18, O.D. × H 22.5 mm × 46 mm, pkg of 100 ea

Bouchon magnétique à visser sur flacons pour espace de tête, filetage de 18 mm

PTFE/silicone septum (white PTFE/blue silicone), septum thickness 1.5 mm, pkg of 100 ea

Molded Thermogreen® LB-2 Septa, with injection hole

diam. 11 mm, pkg of 50 ea

Étalon

Réf. du produit
Description
Tarif

C7 - C40 Saturated Alkanes

certified reference material, 1000 μg/mL each component in hexane

CONDITIONS

sample/matrix

ground Cascade variety hop pellets

SPME fiber

Divinylbenzene/Carboxen/Polydimethylsiloxane (DVB/CAR/PDMS), 50/30 μm (57298-U)

extraction

20 min, headspace, 40 °C

desorption process

3 min, 270 °C

sample preparation

30 min equilibration at 40 °C prior to extraction

sample preparation

fiber postbake after extraction, 3 min at 270 °C

column

Equity-1, 60 m x 0.25 mm I.D., 0.25 μm (28047-U)

oven

60 °C (2 min), 5 °C/min to 275 °C (5 min)

inj. temp.

270 °C

detector

MSD

MSD interface

300 °C

scan range

full scan, m/z 50-500

carrier gas

helium, 1 mL/min constant flow

liner

0.75 mm ID, SPME

Description

Remarque sur l'analyse

Terpenes are contained in the essential oils derived from plants, and often impart characteristic aromas to the plant and/or oil. In this application, headspace-SPME combined with GC/MS was used to analyze some of the terpenes present in the Cascade variety of hops. Chromatographic separation was performed on an Equity-1 capillary GC column, and identification was done using retention indices and spectral library match. Analysis showed a characteristic terpene profile, with high levels of ß-myrcene, caryophyllene, and humulene present.

Autres remarques

add 0.5 gm ground hops to a 10 mL headspace vial

Informations légales