Accéder au contenu
MilliporeSigma

HPLC Analysis of AQC-Amphetamine Enantiomers on Astec® CYCLOBOND I 2000 DMP

HPLC Analysis of AQC-Amphetamine Enantiomers on Astec® CYCLOBOND I 2000 DMP application for HPLC

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

Astec® Cyclobond I 2000 DMP Chiral HPLC Column

5 μm particle size, L × I.D. 25 cm × 4.6 mm

Étalon

Réf. du produit
Description
Tarif

D-Amphetamine hemisulfate salt solution

1.0 mg/mL in methanol, analytical standard, for drug analysis

CONDITIONS

column

CYCLOBOND I 2000 DMP, 25 cm x 4.6 mm I.D., 5 μm particles (20724AST)

mobile phase

[A] 10 mM ammonium acetate, pH 4.1; [B] acetonitrile; (80:20, A:B)

flow rate

1 mL/min

column temp.

30 °C

detector

UV, 220 nm

injection

5 μL

sample

1 mg/mL in methanol