Accéder au contenu
MilliporeSigma

HPLC Analysis of Antibiotics on Ascentis® C18

HPLC Analysis of Antibiotics on Ascentis® C18 application for HPLC

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

Colonne Ascentis® C18 pour HPLC

5 μm particle size, L × I.D. 15 cm × 4.6 mm

CONDITIONS

column

Ascentis C18, 15 cm x 4.6 mm I.D., 5 μm particles (581324-U)

mobile phase

[A] 10 mM monobasic potassium phosphate (pH 3.0 with phosphoric acid); [B] acetonitrile (93:7, A:B)

flow rate

1 mL/min

column temp.

35 °C

detector

UV, 254 nm

injection

10 μL

sample

50 μg/mL each in mobile phase

Description

Informations légales

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany