Misleading results from immunoassays of serum free thyroxine in the presence of rheumatoid factor.

Clinical chemistry (1997-06-01)
A G Norden, R A Jackson, L E Norden, A J Griffin, M A Barnes, J A Little
RESUMO

A novel interference with measurements of serum free thyroxine (FT4) caused by rheumatoid factor (RhF) is described. We found misleading, sometimes gross, increases of FT4 results in 5 clinically euthyroid elderly female patients with high RhF concentrations. All 5 patients had high FT4 on Abbott AxSYM or IMx analyzers. "NETRIA" immunoassays gave misleading results in 4 of the 5 patients; Amerlex-MAB in 2 of 4 patients; AutoDELFIA in 2 of the 5; and Corning ACS-180 and Bayer Diagnostics Immuno 1 in 1 of the 5. BM-ES700 system results for FT4 in these women remained within the reference range. Results for serum T4, thyroid-stimulating hormone, free triiodothyronine, thyroid-hormone-binding globulin, and FT4 measured by equilibrium dialysis were normal in all 5 patients. Drugs, albumin-binding variants, and anti-thyroid-hormone antibodies were excluded as interferences. Addition to normal serum of the RhF isolated from each of the 5 patients increased the apparent FT4 (Abbott AxSYM). Screening of 83 unselected patients demonstrated a highly significant positive correlation between FT4 (Abbott AxSYM) and RhF concentrations. Discrepant, apparently increased FT4 with a normal result for thyroid-stimulating hormone should lead to measurement of the patient's RhF concentration.

MATERIAIS
Número do produto
Marca
Descrição do produto

Sigma-Aldrich
Human IgG−Agarose, suspension