Accéder au contenu
Merck

GC Analysis of FAMEs in Atlantic Salmon on SLB®-IL60

GC Analysis of FAMEs in Atlantic Salmon on SLB®-IL60 suitable for GC

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

SLB®-IL60 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.20 μm

CONDITIONS

column

SLB-IL60, 30 m x 0.25 mm I.D., 0.20 μm (29505-U)

oven

170 °C, 1 °C/min to 225 °C

inj. temp.

250 °C

detector

FID, 250 °C

carrier gas

helium, 1.2 mL/min

injection

1 μL 100:1 split

liner

4 mm I.D., split/splitless type, wool packed single taper FocusLiner design

sample

extract of a farm raised Atlantic salmon, the antioxidant butylated hydroxytoluene (BHT) added prior to extraction

Description

Informations légales

FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany