HPLC Analysis of Phenyl Urea Pesticides on Ascentis® Express C18 Phase
Materialien
Analytische Säule
Produkt-Nr.
Beschreibung
Preisangaben
CONDITIONS
column
Ascentis® Express C18, 10 cm x 4.6 mm I.D., 2.7 μm particles (53827-U)
mobile phase
[A] 0.025 M potassium phosphate buffer, adj. to pH = 2.5, [B] acetonitrile; (50:50, A:B)
flow rate
2.0 mL/min
pressure
300 bar
column temp.
30 °C
detector
UV 245 nm, VWD
injection
0.5 μL
sample
acetonitrile
Beschreibung
Allgemeine Beschreibung
This separation illustrates the use of the highly efficient 2.7 um Ascentis Express Fused-Core C18 stationary phase in the analysis of some popular herbicides. The short run times allow analyses using isocratic conditions so that column equilibration time is not required between runs.