GC Analysis of Water in Acetone (0.5 mg/mL Standard) on Watercol 1910

GC Analysis of Water in Acetone (0.5 mg/mL Standard) on Watercol™ 1910 suitable for GC

Materialien

Analytische Säule

Produkt-Nr.
Beschreibung
Preisangaben

Watercol 1910 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.20 μm

CONDITIONS

column

Watercol 1910, 30 m x 0.25 mm I.D., 0.20 μm (29711-U)

oven

35 °C (4 min), 8 °C/min to 130 °C (1 min)

inj. temp.

250 °C

detector

TCD, 200 °

carrier gas

helium, 11.40 psi constant pressure (equal to 25 cm/sec), measured with oven at 35 °C

injection

1 μL, 100:1 split

liner

4 mm I.D., split/splitless type, wool packed single taper FocusLiner design (2879925-U)

sample

0.5 mg deionized water weighed into a 1 mL volumetric flask, acetone added to the 1 mL line, vortexed, transferred to an autosampler vial

Beschreibung

Hinweis zur Analyse

The "Watercol" series of ionic liquid capillary GC columns are characterized by their ability to produce a sharp peak shape for water, sharp enough so that water can be integrated and quantified. This application shows a calibration standard (0.5 mg/mL water in acetone) on Watercol 1910.