Direkt zum Inhalt
Merck

KVVLA3STT1

Millipore

Durapore®, 0,1 µm, Opticap®-XLT-Kapsel

Opticap® XLT 30, cartridge nominal length 30 in. (15 cm), filter diam. 5.7 in. (14.5 cm), pore size 0.1 μm

Anmeldenzur Ansicht organisationsspezifischer und vertraglich vereinbarter Preise


About This Item

UNSPSC-Code:
23151806

Materialien

polypropylene device
polypropylene housing
polypropylene support
polypropylene vent cap

Qualitätsniveau

Einhaltung gesetzlicher Vorschriften

meets FDA 21 CFR 177-182 (all component materials)

Sterilität

non-sterile

Kompatibilität mit Sterilisationsverfahren

autoclavable compatible

Produktlinie

Opticap® XLT 30

Leistungsmerkmale

hydrophilic

Hersteller/Markenname

Opticap®

Parameter

max. inlet temp. at 25 °C
≤60 mL/min air diffusion at 3.9 bar (56 psig) (in water)
1.7 bar max. differential pressure (25 psid) at 80 °C (Forward)
2.8 bar max. inlet pressure (40 psi) at 60 °C
3.5 bar max. differential pressure (50 psid) at 25 °C (Reverse; intermittent)
5.5 bar max. differential pressure (80 psid) at 25 °C (Forward)
5.5 bar max. inlet pressure (80 psi) at 23 °C
80 psig max. inlet pressure

Methode(n)

sterile filtration: suitable

Länge

34.3 in.

Breite

6.0 in.

Nennlänge Kartusche

30 in. (15 cm)

device size

30 in.

Filterdurchm.

5.7 in. (14.5 cm)

filter filtration area

2.1 m2

inlet to outlet width

34.3 in. (87.1 cm)

Verunreinigungen

≤0.5 EU/mL bacterial endotoxins (LAL test, aqueous extraction)
<0.5 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)

Matrix

Durapore®

Porengröße

0.1 μm pore size

Blasenpunkt

≥4830 mbar (70 psig), air with water at 23 °C

Fitting

1/4 in. drain/vent connection (with double O-ring Seal)
(38 mm (1 1/2 in.) Sanitary Flange Inlet and Outlet)

Suchen Sie nach ähnlichen Produkten? Aufrufen Leitfaden zum Produktvergleich

Allgemeine Beschreibung

Device Configuration: Cartridge
Cartridge Type: Liquid

Leistungsmerkmale und Vorteile

T-line/Gauge Port: T-line

Angaben zur Herstellung

Sterilization Method: Capsule integrity was maintained after 3 autoclave cyclces of 60 minutes at 126 °C.
This product was manufactured with a Durapore® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).

Hinweis zur Analyse

Bacterial Retention: Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 107 CFU/cm2 using ASTM® F838 methodology.
Gravimetric Extractables: ≤ 75 mg per capsule after 24 hours in water at controlled room temperature.

Sonstige Hinweise

Directions for Use:
  • Organism Retention: Microorganism
  • Mode of Action: Filtration (size exclusion)
  • Application: BioProcessing
  • Intended Use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer Wetting Instructions for Filter Units with Durapore® Membrane user guide
  • Storage Statement: Store in dry location
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations

These products are manufactured in a facility which adheres to FDA Good Manufacturing Practices.

Rechtliche Hinweise

ASTM is a registered trademark of American Society for Testing and Materials
Durapore is a registered trademark of Merck KGaA, Darmstadt, Germany
OPTICAP is a registered trademark of Merck KGaA, Darmstadt, Germany

Not finding the right product?  

Try our Produkt-Auswahlhilfe.


Analysenzertifikate (COA)

Suchen Sie nach Analysenzertifikate (COA), indem Sie die Lot-/Chargennummer des Produkts eingeben. Lot- und Chargennummern sind auf dem Produktetikett hinter den Wörtern ‘Lot’ oder ‘Batch’ (Lot oder Charge) zu finden.

Besitzen Sie dieses Produkt bereits?

In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.

Die Dokumentenbibliothek aufrufen

Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..

Setzen Sie sich mit dem technischen Dienst in Verbindung.