HPLC Analysis of Aflatoxins on Ascentis® Express C18

HPLC Analysis of Aflatoxins on Ascentis® Express C18 application for HPLC

Materiales

analytical column

Referencia del producto
Descripción
Precios

Columna para HPLC Ascentis® Express C18, 2,7 μm

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

standard

Referencia del producto
Descripción
Precios

Aflatoxin Mix 4 solution

analytical standard

CONDITIONS

column

Ascentis Express C18, 10 cm x 2.1 mm I.D., 2.7 μm particles (53823-U)

mobile phase

[A] water; [B] methanol; [C] acetonitrile; (720:120:120, A:B:C) containing 0.780 g/L potassium bromide, 230 μL/L nitric acid

flow rate

0.400 mL/min

pressure

4568 psi (315 bar)

column temp.

35 °C

detector

FLD excitation = 360 nm, emission = 440 nm

injection

20 μL

sample

200 μL spiked 84:16 acetonitrile:di water in 880 μL di water, Spike concentrations: B1 = 16 ppb; B2 = 4 ppb; G1 = 16 ppb; B2 = 4 ppb

Descripción

Analysis Note

Aflatoxins are a complex of closely related toxins produced as secondary metablolites of fungi. They are found in grains and nuts such as corn, wheat, or peanuts. They are produced by aspergillus flavus and a. Parasiticus. Members of the group include aflatoxin b1, aflatoxin b2, aflatoxin g1, aflatoxin g2, aflatoxin m1, and aflatoxin m2.