US EPA Method 625: GC Analysis of Semivolatiles on SLB®-5ms
Materiales
analytical column
For use with
Referencia del producto
Descripción
Precios
Wheaton wide-mouth bottle
amber soda-lime glass bottle, capacity 1000 mL, white polypropylene cap, (PTFE faced LDPE liner), case of 12 eaguard column
Referencia del producto
Descripción
Precios
standard
Referencia del producto
Descripción
Precios
EPA 8270 GC-MS Tuning Solution II
certified reference material, 1000 μg/mL each component in dichloromethane625 Semivolatile Calibration Mix
1000 μg/mL each component in methylene chloride: benzene (3:1), analytical standardCLP Semivolatiles Surrogate Standards Mix
certified reference material, 4000 μg/mL each component in methylene chlorideSemivolatile Internal Standards Mix
certified reference material, TraceCERT®, 2000 μg/mL each component in methylene chloride, ampule of 2 × 1 mLCONDITIONS
column
SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (28471-U)
oven
40 °C (2 min), 22 °C/min to 240 °C, 10 °C/min to 330 °C (5 min)
inj. temp.
250 °C
MSD interface
330 °C
scan range
m/z = 40-450
carrier gas
helium, 1 mL/min constant (10 min), ramped to 1.5 mL/min, hold for remainder of run
injection
1 μL, pulsed (15 psi until 0.10 min) splitless (0.50 min)
liner
4 mm I.D., single taper
sample
79 component semivolatile standard (each at 50 ppm), plus 6 internal standards (each at 40 ppm) in methylene chloride
Descripción
Legal Information