Open the foil pouch, dissolve contents in 900 mL of water; add additional water to a final volume of 1 liter to make a 0.1 M solution at a temperature of 25 °C. No pH adjustment is necessary. Guaranteed accuracy: ±0.1 pH units.
Tris has a significant temperature coefficient, so select the mixture to give the desired final pH at the desired temperature. From 5 °C to 25 °C, the pH decreases an average of 0.03 pH units per °C. From 25 °C to 37 °C, the pH decreases an average of 0.025 pH units per °C.
If a different concentration is desired, note that there is a concentration effect on the pH of the solution.
The product is cell culture tested. It is also tested for low endotoxin levels and analyzed for the absence of nucleases and proteases.
Packaging
In convenient foil pouches
Other Notes
This product is a pre-mixed combination of Trizma base and Trizma HCl in convenient foil pouches that can be used to prepare a common pH value for Tris buffer.
Certain pH electrodes do not give accurate results with Tris buffers.
Preparation Note
Trizma solutions can be autoclaved.
Legal Information
Trizma is a registered trademark of Merck KGaA, Darmstadt, Germany
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.