Accéder au contenu
Merck

HPLC Analysis of Dopa and Carbidopa Enantiomers on Astec® CHIROBIOTIC® T

HPLC Analysis of Dopa and Carbidopa Enantiomers on Astec® CHIROBIOTIC® T application for HPLC

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

Colonne chirale pour HPLC Astec® CHIROBIOTIC® T

5 μm particle size, L × I.D. 25 cm × 4.6 mm

Étalon

Réf. du produit
Description
Tarif

S-(−)-Carbidopa

≥98% (TLC), powder

3,4-Dihydroxy-D-phenylalanine

powder, ≥95%

3,4-Dihydroxy-DL-phenylalanine

3,4-Dihydroxy-L-phenylalanine

≥98% (TLC)

CONDITIONS

column

CHIROBIOTIC T, 25 cm x 4.6 mm I.D., 5 μm particles (12024AST)

mobile phase

[A] 20 mM ammonium acetate, pH 4; [B] methanol; (30:70, A:B)

flow rate

1 mL/min

column temp.

25 °C

detector

UV, 230 nm

injection

5 μL

sample

1 mg/mL in methanol

Description

Informations légales

Astec is a registered trademark of Merck KGaA, Darmstadt, Germany
CHIROBIOTIC is a registered trademark of Sigma-Aldrich Co. LLC