G005907
HPLC Analysis of Aflatoxins in Wheat on Ascentis® Express C18 after SPE using Supel™ Tox AflaZea
application for HPLC
About This Item
technique(s)
HPLC: suitable
test parameters
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: 25 g wheat meal spiked with pesticide standards
SPE tube/cartridge: Supel™Tox AflaZea, 6 mL (55314-U)
sample addition: 2 mL
elution: collect eluate in 5 mL silanized culture tubes
eluate post-treatment: combine 200 μL of sample eluate and 880 μL of DI water in a 2 mL autosampler vial
column: Ascentis Express C18, 10 cm x 2.1 mm I.D., 2.7 μm particles (53823-U)
mobile phase: [A] water; [B] methanol; [C] acetonitrile; (720:120:120, A:B:C) containing 0.780 g/L potassium bromide, 230 μL/L nitric acid
flow rate: 0.400 mL/min
pressure: 4568 psi (315 bar)
column temp.: 35 °C
detector: FLD 360 ex; 440 em
injection: 20 μL
derivatization: via KOBRA electrochemical cell, 50 μA
suitability
application for HPLC
application(s)
food and beverages
Analysis Note
Weigh 25 grams of wheat meal into a 500 mL Erlenmeyer flask. Add 100 mL of 84:16 acetonitrile:water, seal, and place on an orbital shaker at med-high speed for 90 minutes. Filter the sample through a #2 qualitative filter. Spike sample at 2 mg/mL using analytical standards.
Legal Information
Analito
Analytical column
Standard
related product
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
Lo sentimos, en este momento no disponemos de COAs para este producto en línea.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico