HPLC Analysis of Steroids on Ascentis® Express C18 (10 cm x 2.1 mm I.D.)

application for HPLC

Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización


About This Item

technique(s)

HPLC: suitable

test parameters

column: Ascentis Express C18, 10 cm x 2.1 mm I.D., 2.7 μm particles (53823-U)
mobile phase: [A] water; [B] acetonitrile; (60:40, A:B)
flow rate: 0.2 mL/min
pressure: 2944 psi (203 bar)
column temp.: 35 °C
detector: UV, 254 nm
injection: 2 μL
sample: 50 mg/L in 75:25, water:methanol

suitability

application for HPLC

application(s)

clinical

Analysis Note

This application demonstrates the suitability of Ascentis Express C18 for the efficient separation of 5 steroids.