G1006858
LC/MS Analysis of Nucleosides on Ascentis® Express F5
application for LC-MS
About This Item
Productos recomendados
technique(s)
LC/MS: suitable
test parameters
column: Ascentis Express F5, 15 cm x 2.1 mm I.D., 2.7 μm particles, 90Å (53571-U)
mobile phase: [A] 5 mM ammonium acetate, pH 5.5 with acidic acid in water; [B] Methanol
gradient: hold at 0% for 6 min; to 50% B in 4 min; hold at 50% B for 4 min; to 0% B in 0.1 min; to hold at 0% B for 5.9 min
flow rate: 0.3 mL/min
pressure: 5017-9367 psi
column temp.: 15 °C
detector: MS, ESI (-) QTOF, EIC
injection: 5 μL
sample: 100 μg/mL of 12 nucleosides except N6-methyladenosine in 99:1 10mM ammonium acetate pH 5.5 in water : methanol
suitability
application for LC-MS
application(s)
life science and biopharma
General description
Analysis Note
Other Notes
Legal Information
Analito
analytical column
mobile phase component
standard
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
Lo sentimos, en este momento no disponemos de COAs para este producto en línea.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico