G006299
LC/MS Analysis of Opioid Glucuronide Metabolites in Urine on Ascentis® Express F5 after Solid Phase Extraction (SPE) using Supel™-Select HLB
application for SPE, application for HPLC
About This Item
technique(s)
HPLC: suitable
solid phase extraction (SPE): suitable
test parameters
sample preparation: Solid Phase Extraction
sample/matrix: urine spiked with drugs and metabolites at 300 ng/mL
SPE well plate: Supel-Select HLB 96-well plate, 60 mg/well (575662-U)
condition: 1 mL methanol followed by 1 mL water
sample addition: 1 mL hydrolyzed urine
washing: 1 mL water followed by 1 mL 25% methanol
elution: 1 mL acetonitrile
column: Ascentis Express F5, 10 cm x 2.1 mm I.D., 2.7 μm particles (53569-U)
mobile phase: [A] 5 mM ammonium formate, pH 4.2 (adjusted with formic acid); [B] 5 mM ammonium formate in (95:5 acetonitrile:water) pH 5.2 (adjusted with formic acid)
gradient: 15% B held for 1 min, to 70% B in 2 min, held at 70% B for 7 min
flow rate: 0.4 mL/min
column temp.: 35 °C
detector: ESI(+) TOF, full scan, m/z 100-1000
injection: 3 μL
sample: 300 ng/mL
suitability
application for HPLC
application for SPE
application(s)
clinical
forensics and toxicology
Analysis Note
Legal Information
Analito
SPE tube or plate
analytical column
standard
Choose from one of the most recent versions:
Certificados de análisis (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico