Saltar al contenido
MilliporeSigma
Todas las fotos(1)

Documentos clave

G005755

GC Analysis of Basic Drugs on SLB®-5ms (20 m x 0.18 mm I.D., 0.36 μm) after SPME using 65 μm PDMS/DVB Fiber, Fast GC Analysis

suitable for GC, application for SPME

Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización


About This Item

technique(s)

gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable

test parameters

sample/matrix: 500 μg/L of each compound in 2 mL 0.05M phosphate buffer, 25% sodium chloride, pH=11
SPME fiber: Polyacrylate, 85 μm (57304); PDMS/DVB Stableflex, 65 μm (57326-U)
extraction: 15 min, fiber immersion in water, room temperture, stirring at 450 rpm
desorption process: 2 min, 250 °C
column: SLB-5ms, 20 m x 0.18 mm I.D., 0.36 μm (28576-U)
oven: 100 °C (1 min), 20 °C/min to 325 °C (5 min)
detector: MSD, m/z 40 - 450
MSD interface: 325 °C
carrier gas: helium, 1.5 mL/min, constant
liner: 0.75 mm I.D. SPME

suitability

application for SPME

application(s)

forensics and toxicology

Legal Information

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany

Analito

Descripción

    4-Chloro-DL-phenylalanine
    Codeine
    Diazepam
    Heroin

Elija entre una de las versiones más recientes:

Certificados de análisis (COA)

Lot/Batch Number

Lo sentimos, en este momento no disponemos de COAs para este producto en línea.

Si necesita más asistencia, póngase en contacto con Atención al cliente

¿Ya tiene este producto?

Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.

Visite la Librería de documentos

Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.

Póngase en contacto con el Servicio técnico