Saltar al contenido
MilliporeSigma
Todas las fotos(1)

Documentos

G005589

HPLC Analysis of Prednisolone (Secondary Pharma Standard) on Ascentis® C18

application for HPLC

Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización


About This Item

technique(s)

HPLC: suitable

test parameters

column: Ascentis C18, 15 cm x 4.6 mm I.D., 5 μm particles (581324-U)
mobile phase: [A] water; [B] methanol:acetonitrile (50:50, A:B)
gradient: 40% B for 14 min; to 80% B in 6 min; held at 80% B for 5 min
flow rate: 1 mL/min
column temp.: 40 °C
detector: UV, 254 nm
injection: 10 μL
sample: 125 μg/mL in water:acetonitrile (60:40)

suitability

application for HPLC

application(s)

pharmaceutical (small molecule)

Legal Information

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analito

Descripción

    Prednisolone, pharmaceutical secondary standard; traceable to USP, PhEur and BP

analytical column

Referencia del producto
Descripción
Precios

Choose from one of the most recent versions:

Certificados de análisis (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Atención al cliente

¿Ya tiene este producto?

Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.

Visite la Librería de documentos

Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.

Póngase en contacto con el Servicio técnico