G005295

GC Analysis of Trichloroanisole (TCA) and Precursors in Red Wine (Spiked) on SLB-5ms after SPME using 100 μm PDMS Fiber

suitable for GC, application for SPME

Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización


About This Item

technique(s)

gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable

test parameters

sample/matrix: 1.5 mL red wine spiked with each analyte at 100 ng/L + 600 μL 5% potassium carbonate (K2CO3) + 240 mg sodium chloride (NaCl) + 60 μL acetic anhydride
SPME fiber: metal fiber coated with 100 μm PDMS (57928-U)
extraction: headspace, 50 °C for 30 min, with stirring
desorption process: 250 °C for 3 min
column: SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (28471-U)
oven: 50 °C (1 min), 25 °C/min to 280 °C
detector: ECD, 290 °C
carrier gas: helium, 1.5 mL/min, constant
liner: 0.75 mm I.D. SPME

suitability

application for SPME

application(s)

food and beverages

Analytical column

Referencia del producto
Descripción
Precios

related product

Referencia del producto
Descripción
Precios

standard

Referencia del producto
Descripción
Precios

Elija entre una de las versiones más recientes:

Certificados de análisis (COA)

Lot/Batch Number

Lo sentimos, en este momento no disponemos de COAs para este producto en línea.

Si necesita más asistencia, póngase en contacto con Atención al cliente

¿Ya tiene este producto?

Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.

Visite la Librería de documentos

Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.

Póngase en contacto con el Servicio técnico