794-0399
GC Analysis of a 20-Component Organochlorine Pesticide Mix in Water on a 5% Phenyl Column after SPE using Supelclean™ ENVI™-8
suitable for GC, application for SPE
Sinónimos:
Chlorinated pesticides in hazardous waste
About This Item
technique(s)
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
test parameters
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: waste water spiked with pesticides
SPE tube/cartridge: Supelclean ENVI-8, glass hardware, 500 mg/6 mL (57107 (500 mg/6 mL); 57108 (1 g/6 mL))
condition: 3 mL methanol; h 2 mL 5% methanol in water
sample addition: load sample at 5 mL/min
drying: 2-3 min with nitrogen stream
elution: 2 x 4 mL hexane:acetone (90:10), drop wise. allow to soak into bed
column: 5% Phenyl Polysiloxane, 15 m × 0.53 mm I.D., 0.50 μm (Supelco equivalent, Equity-5, 28252-U)
oven: 150 °C (0.5 min) to 275 °C (5 min)
inj. temp.: 200 °C
detector: ECD, 300 °C
carrier gas: helium
injection: 1 μL split/splitless (45 sec delay)
suitability
application for SPE
application(s)
environmental
food and beverages
Application
Add 5 mL methanol to 100 mL aqueous hazardous waste. Adjust to pH 5-7 if necessary.
Legal Information
Analito
SPE tube or plate
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documentos section.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico