TLC Analysis of Herbicides / Fungicides on HPTLC DIOL F254s
Materials
accessory
Product No.
Description
Pricing
Hamilton® TLC syringes
1702, volume 25 μL, needle size 22s ga, needle L 51 mm (2 in.),coated tip
Hamilton® TLC syringes
1705, volume 50 μL, needle size 22 ga, needle L 51 mm (2 in.),coated tip
Hamilton® TLC syringes
1701, volume 10 μL, needle size 26s ga, needle type, fixed, needle L 51 mm (2 in.),coated tip
Hamilton® TLC syringes
1710, volume 100 μL, needle size 22 ga, needle L 51 mm (2 in.),coated tip
CONDITIONS
plate
1.05636 HPTLC pre-coated plate DIOL F 254s (1.05636.0001, 1.12668.0001 )
eluent
[A] petroleum benzine (40 - 60°); [B] acetone; (85:15, A:B)
migration distance
7 cm
chamber
normal chamber without chamber saturation
sample
0.5 μL
detector
in situ evaluation with TLC/HPTLC scanner (Camag) UV 238 nm