- Ultrasonic extraction with ultra-performance liquid chromatography/tandem mass spectrometry for the determination of ochratoxin A in processed cereal products.
Ultrasonic extraction with ultra-performance liquid chromatography/tandem mass spectrometry for the determination of ochratoxin A in processed cereal products.
Journal of AOAC International (2015-04-24)
Kieu Thi Ngoc Nguyen, Dojin Ryu
PMID25902987
ABSTRACT
A rapid, simple, and reliable method using ultra-performance LC/MS/MS (UPLC/MS/MS) was developed for determination of ochratoxin A (OTA) in processed cereal products. OTA was ultrasonically extracted from the sample with acetonitrile-water (80 + 20, v/v), and the extract was then injected into the UPLC/MS/MS system after filtration. The calibration curves had good linearity with coefficients of determination greater than 0.999. Recoveries of OTA were in the range of 90-104%. LOD and LOQ of OTA in samples were 0.6 and 2.0 ng/g, respectively, and no significant matrix effect was found. This method was applied to determine OTA in 25 oat-based cereal samples. OTA was detected in five samples (20%) in the range of 2.4 to 7.3 ng/g.
MATERIALS
Product Number
Brand
Product Description
Sigma-Aldrich
Acetonitrile solution, contains 0.1 % (v/v) trifluoroacetic acid, suitable for HPLC
Sigma-Aldrich
Acetonitrile solution, contains 10.0% acetone, 0.05% formic acid, 40.0% 2-propanol
Sigma-Aldrich
Acetonitrile solution, contains 0.05 % (w/v) ammonium formate, 5 % (v/v) water, 0.1 % (v/v) formic acid, suitable for HPLC
Supelco
Residual Solvent - Acetonitrile, Pharmaceutical Secondary Standard; Certified Reference Material
Sigma-Aldrich
Acetic acid, glacial, puriss., meets analytical specification of Ph. Eur., BP, USP, FCC, 99.8-100.5%
Sigma-Aldrich
Acetic acid, glacial, puriss. p.a., ACS reagent, reag. ISO, reag. Ph. Eur., ≥99.8%