  • Simultaneous analysis of five antidepressant drugs using direct injection of biofluids in a capillary restricted-access media-liquid chromatography-tandem mass spectrometry system.

Simultaneous analysis of five antidepressant drugs using direct injection of biofluids in a capillary restricted-access media-liquid chromatography-tandem mass spectrometry system.

Journal of chromatography. A (2008-03-22)
Alvaro J Santos-Neto, Jonas Bergquist, Fernando M Lanças, Per J R Sjöberg
ABSTRACT

Direct analysis, with minimal sample pretreatment, of antidepressant drugs, fluoxetine, imipramine, desipramine, amitriptyline, and nortriptyline in biofluids was developed with a total run time of 8 min. The setup consists of two HPLC pumps, injection valve, capillary RAM-ADS-C18 pre-column and a capillary analytical C18 column connected by means of a six-port valve in backflush mode. Detection was performed with ESI-MS/MS and only 1 microm of sample was injected. Validation was adequately carried out using FLU-d(5) as internal standard. Calibration curves were constructed under a linear range of 1-250 ng mL(-1) in plasma, being the limit of quantification (LOQ), determined as 1 ng mL(-1), for all the analytes. With the described approach it was possible to reach a quantified mass sensitivity of 0.3 pg for each analyte (equivalent to 1.1-1.3 fmol), translating to a lower sample consumption (in the order of 10(3) less sample than using conventional methods).

MATERIALS
Product Number
Brand
Product Description

Ketobemidone impurity C, European Pharmacopoeia (EP) Reference Standard
Pricing and availability is not currently available.