跳轉至內容
Merck
  • Determination of nitrite in waters by microplate fluorescence spectroscopy and HPLC with fluorescence detection.

Determination of nitrite in waters by microplate fluorescence spectroscopy and HPLC with fluorescence detection.

Analytical chemistry (1999-08-18)
A Büldt, U Karst
摘要

A selective and versatile fluorescence spectroscopic method for the determination of nitrite in waters has been developed. Nitrite reacts in the presence of mineral acids with the nonfluorescent N-methyl-4-hydrazino-7-nitrobenzofurazan forming N-methyl-4-amino-7-nitrobenzofurazan, which can be detected by fluorescence spectroscopy with an excitation maximum at lambda = 468 nm and an emission maximum at lambda = 537 nm in acetonitrile. Three new methods based on this reaction have been developed: Direct fluorescence spectroscopy, HPLC/fluorescence, or HPLC with UV/vis detector may be selected as detection techniques. On microplates, high-throughput fluorescence spectroscopy is achieved, while HPLC/fluorescence provides lower limits of detection, and HPLC with UV/vis detection enables evaluation of the reaction with standard instrumentation. Different water samples were investigated using all detection modes, and a photometric standard procedure was successfully employed to validate the new methods with an independent technique.