Skip to Content
Merck

GC Analysis of Trichloroanisole (TCA) and Precursors in Red Wine on SLB®-5ms after SPME using 100 μm PDMS Fiber

GC Analysis of Trichloroanisole (TCA) and Precursors in Red Wine on SLB®-5ms after SPME using 100 μm PDMS Fiber suitable for GC, application for SPME

Materials

analytical column

Product No.
Description
Pricing

SLB®-5ms Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

SPME fiber

Product No.
Description
Pricing

SPME fiber assembly Polydimethylsiloxane (PDMS)

metal alloy fiber, df 100 μm(PDMS, needle size 23 ga, for use with autosampler

CONDITIONS

sample/matrix

1.5 mL red wine + 600 μL 5% potassium carbonate + 240 mg sodium chloride + 60 μL acetic anhydride

SPME fiber

metal fiber coated with 100 μm PDMS (57928-U)

extraction

headspace, 50 °C for 30 min., with stirring

desorption process

250 °C for 3 min

column

SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (28471-U)

oven

50 °C (1 min), 25 °C/min to 280 °C

detector

ECD, 290 °C

carrier gas

helium, 1.5 mL/min, constant

liner

0.75 mm I.D. SPME

Description

Legal Information

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany