G005743

HPLC Analysis of Steroids on Ascentis® Express F5

application for HPLC

Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización


About This Item

technique(s)

HPLC: suitable

test parameters

column: Ascentis Express F5, 15 cm x 4.6 mm I.D., 2.7 μm particles (53591-U)
mobile phase: [A] water; [B] methanol; (50:50, A:B)
flow rate: 0.8 mL/min
pressure: 2393 psi (165 bar)
column temp.: 35 °C
detector: UV, 240 nm
injection: 5 μL
sample: 10 μg/mL of each in 90: 10, water: methanol

suitability

application for HPLC

application(s)

clinical

Analysis Note

Sample concentrations at 10 ug/mL in 90:10 water:methanol. Acetonitrile does not provide adequate selectivity on this phase. Mass spec was used for peak identification.

Legal Information

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analito

Descripción

    Hydrocortisone
    Prednisolone
    Prednisone

analytical column

Referencia del producto
Descripción
Precios

standard

Referencia del producto
Descripción
Precios

Elija entre una de las versiones más recientes:

Certificados de análisis (COA)

Lot/Batch Number

Lo sentimos, en este momento no disponemos de COAs para este producto en línea.

Si necesita más asistencia, póngase en contacto con Atención al cliente

¿Ya tiene este producto?

Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.

Visite la Librería de documentos

Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.

Póngase en contacto con el Servicio técnico