G004845
GC Analysis of Resveratrol in Red Wine (Spiked) on SLB®-5ms after SPME using a 85 μm Polyacrylate Fiber
suitable for GC, application for SPME
About This Item
técnicas
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
parámetros analíticos
sample/matrix: 3 mL of red wine (California merlot) spiked with resveratrol at 100 μg/L then diluted 3:1 in 12% ethanol:water
SPME fiber: 85 μm polyacrylate (57304, 57305, 57294-U)
extraction: immersion for 15 min at room temperature while stirring at 400 rpm, fiber gently blotted with a Kimwipe to remove excess water, (then derivatization by immersion in a 4 mL vial containing 5 μL of Sylon-BFT (fresh vial for each analysis than was allowed to equilabrate 60-90 minutes prior to use) for 20 min)
desorption process: 280 °C for 2 min
column: SLB-5ms; 30 m x 0.25 mm I.D., 0.25 μm (28471-U)
oven: 100 °C (1 min), 10 °C/min to 325 °C (3 min)
inj. temp.: 280 °C
detector: GC/MS
MSD interface: 325 °C
scan range: m/z 40-450
carrier gas: helium, 1 mL/min, constant
liner: 0.75 mm I.D., SPME
sample: Red wine
idoneidad
application for SPME
aplicaciones
food and beverages
Nota de análisis
Información legal
Estándar
Fibra para SPME
Para utilizar con
Producto relacionado
columna analítica
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documentos section.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico