跳转至内容
Merck

LC/MS Analysis of Illicit Bath Salts on Ascentis® Express HILIC, Fast Analysis

LC/MS Analysis of Illicit Bath Salts on Ascentis® Express HILIC, Fast Analysis application for HPLC

材料

分析柱

产品编号
说明
价格

Ascentis® Express HILIC, 2.7 μm HPLC色谱柱

2.7 μm particle size, L × I.D. 5 cm × 2.1 mm

標準

产品编号
说明
价格

Butylone hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

Ethylone hydrochloride

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

布洛芬标准液 盐酸盐 溶液

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

Methylone hydrochloride

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

Methedrone hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

column

Ascentis Express HILIC, 5 cm x 2.1 mm I.D., 2.7 μm particles (53934-U)

mobile phase

5 mM ammonium formate in acetonitrile:water (95:5)

flow rate

0.6 mL/min

column temp.

35 °C

detector

ESI(+), TIC 100 - 1000 m/z

sample

1 μL

说明

分析報告

With the rapid development of unregulated designer and synthetic compounds, the field of illicit drug testing has recently been met with a changing environment. Of most concern has been the development of a class of phenethylamine and cathinone compounds being marketed as “bath salts,” “jewelry cleaner,” or “plant food.” Forensic testing facilities often experience difficulty in testing these compounds due to the fact that they are not detected under normal ELISA testing methods; additional more specific LC/MS methods are necessary. The challenge for LC/MS detection of these particular bath salts resides in three sets of isobaric compounds that require chromatographic resolution for positive confirmation. For example, both butylone and ethylone have the same mono isotopic mass, making these compounds indistinguishable, even when using accurate mass time of flight TOF-MS. Efficient chromatographic separation is necessary for accurate quantitation of these compounds. The polar basic nature of the bath salts makes these compounds difficult to retain on traditional reversed-phase C18 and even polar embedded stationary phases, making them prime subjects for HILIC chromatographic separation. Shown here is the fast, high-resolution separation of nine synthetic bath salts on Ascentis Express HILIC. Reliable quantitation is achieved by using Cerilliant CRMs.

法律資訊