跳转至内容
Merck

GC Analysis of PCBs in Whole Cow Milk (Spiked) on SLB®-5ms (20 m x 0.18 mm I.D., 0.18 μm) after QuEChERS Cleanup using Supel QuE Z-Sep, Fast GC Analysis

GC Analysis of PCBs in Whole Cow Milk (Spiked) on SLB®-5ms (20 m x 0.18 mm I.D., 0.18 μm) after QuEChERS Cleanup using Supel™ QuE Z-Sep, Fast GC Analysis suitable for GC, application for SPE

材料

SPE管或板

产品编号
说明
价格

Supel QuE

Empty Centrifuge Tube with Lid, centrifuge tube volume 50 mL, pk of 50, suitable for EN 15662:2008 per BS

Supel QuE

Non-Buffered Tube 1, pk of 50

Supel QuE

Z-Sep+ Bulk, pkg of 20 g

分析柱

产品编号
说明
价格

SLB ®-5 ms 毛细管 GC 色谱柱

L × I.D. 20 m × 0.18 mm, df 0.18 μm

標準

产品编号
说明
价格

PCB 28、52、101、138、153、180、209 溶液

analytical standard, 10 μg/mL in isooctane

CONDITIONS

sample preparation

SPE (Solid Phase Extraction)

sample/matrix

10 g whole cow′s milk

extraction tube

Supel QuE Non-buffered tube (4 g magnesium sulfate, 1 g sodium chloride) (55294-U)

clean-up tube

Supel QuE Z-Sep 150 mg, magnesium sulfate (150 mg, part no. 63135) (55299-U)

column

SLB-5ms, 20 m x 0.18 mm I.D., 0.18 μm (28564-U)

oven

75 °C (1 min), 12 °C/min to 340 °C (10 min)

inj. temp.

250 °C

detector

μ-ECD, 340 °C

carrier gas

hydrogen, 1.2 mL/min, constant

injection

1 μL splitless (0.75 min)

liner

4 mm I.D. FocusLiner with taper

说明

一般說明

Reference: Stenerson, K.; Ye, M.; Halpenny, M., New Analytical Tools for the Determination of Persistent Organic Pollutants (POPs) in Fatty Food and Beverage Matrices Using QuEChERS Extraction/Cleanup and Gas Chromatography (GC) Analysis. Pittcon 2013. Sigma-Aldrich/Supelco publication T413034

分析報告

This application demonstrates the use of dispersive SPE or "QuEChERS" for the extraction and cleanup of whole cow′s milk for GC-ECD analysis of polychlorinated biphenyl (PCBs). Z-Sep sorbent was used in the cleanup step and found to reduce background and yield recoveries >70% for all PCB congeners evaluated except no. 209, which was recovered at 68%.
Extraction solvent: acetone:hexane (1:1), 20 mL
Clean-up volume: 1 mL

法律資訊

FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany
Supel is a trademark of Sigma-Aldrich Co. LLC