跳转至内容
Merck

GC Analysis of Class 2 Residual Solvents on Equity-1 after SPME using 85 μm Polyacrylate Fiber

GC Analysis of Class 2 Residual Solvents on Equity-1 after SPME using 85 μm Polyacrylate Fiber suitable for GC

材料

固相微萃取纖維

产品编号
说明
价格

聚丙烯酸酯 (PA) 固相微萃取萃取头

df 85 μm(PA, for use with manual holder, needle size 24 ga

CONDITIONS

sample/matrix

2 mL water + sodium chloride at 25% (pH 11) containing each analyte at 5 ppm, in a 4 mL vial

SPME fiber

85 μm Polyacrylate, 24 ga (57304)

extraction

eated headspace, 60 °C for 5 min

desorption flow

250 °C for 3 min

column

Equity-1, 10 m x 0.20 mm I.D., 1.2 μm (28043-U)

oven

40 °C (0.75 min), 20 °C/min to 200 °C (10 min)

inj. temp.

250 °C

detector

FID

carrier gas

helium, 35 cm/sec @ 40 °

injection

5:1 split

liner

0.75 mm I.D. SPME type

说明

法律資訊