推荐产品
技術
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
測試參數
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: 10 g whole cow′s milk
extraction tube: Supel™ QuE Non-buffered tube (4 g magnesium sulfate, 1 g sodium chloride) (55294-U)
clean-up tube: Supel™ QuE Z-Sep 150 mg, magnesium sulfate (150 mg, part no. 63135) (55299-U)
column: SLB-5ms, 20 m x 0.18 mm I.D., 0.18 μm (28564-U)
oven: 75 °C (1 min), 12 °C/min to 340 °C (10 min)
inj. temp.: 250 °C
detector: μ-ECD, 340 °C
carrier gas: hydrogen, 1.2 mL/min, constant
injection: 1 μL splitless (0.75 min)
liner: 4 mm I.D. FocusLiner™ with taper
適合性
application for SPE
應用
food and beverages
一般說明
Reference: Stenerson, K.; Ye, M.; Halpenny, M., New Analytical Tools for the Determination of Persistent Organic Pollutants (POPs) in Fatty Food and Beverage Matrices Using QuEChERS Extraction/Cleanup and Gas Chromatography (GC) Analysis. Pittcon 2013. Sigma-Aldrich/Supelco publication T413034
分析報告
This application demonstrates the use of dispersive SPE or "QuEChERS" for the extraction and cleanup of whole cow′s milk for GC-ECD analysis of polychlorinated biphenyl (PCBs). Z-Sep sorbent was used in the cleanup step and found to reduce background and yield recoveries >70% for all PCB congeners evaluated except no. 209, which was recovered at 68%.
Extraction solvent: acetone:hexane (1:1), 20 mL
Clean-up volume: 1 mL
Extraction solvent: acetone:hexane (1:1), 20 mL
Clean-up volume: 1 mL
法律資訊
FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany
Supel is a trademark of Sigma-Aldrich Co. LLC
SPE管或板
我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.
联系技术服务部门