跳转至内容
Merck
所有图片(1)

Key Documents

G005790

LC/MS Analysis of L-Methionine Sulfoximine, Sulfone and Sulfoxide on Ascentis® Express HILIC

application for HPLC

登录查看公司和协议定价


About This Item

技術

HPLC: suitable

測試參數

column: Ascentis Express HILIC, 10 cm x 3.0 mm I.D., 2.7 μm particles (53970-U)
mobile phase: [A] acetonitrile; [B] water with 0.1% formic acid, pH 3.5 with ammonium formate; (75:25, A:B)
flow rate: 0.4 mL/min
pressure: 1350 psi 93 bar
temp.: 25 °C
detector: MS, ESI (+), TIC, m/z 150-300
injection: 5 μL
sample: 10 μg/mL in 90:10, methanol:water

適合性

application for HPLC

應用

life science and biopharma

分析報告

L-methionine sulfoximine and L-buthionine sulfoximine are used in are used to prevent additional enzyme activity in Chinese hampster ovary (CHO) cell lines with supplemental Glutamine Synthase (GS). It is of interest to be able to monitor both purity of such molecules in order to control addition to cell media as well as for the assay of parent molecules during use. The analytes are highly polar and thus should be amenable to HILIC-LC-MS analysis. This first installment investigates the ability of Ascentis Express HILIC and Ascentis Express OH5 stationary phases for the separation of the methionine analog and related compounds as well as methionine and buthionine separations.

法律資訊

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

分析物

说明

    L-Methionine sulfone
    L-Methionine sulfoximine, suitable for cell culture
    L-Methionine sulfoxide

从最新的版本中选择一种:

分析证书(COA)

Lot/Batch Number

抱歉,我们目前尚未线上提供该产品的COA。

如需帮助,请联系 客户支持

已有该产品?

在文件库中查找您最近购买产品的文档。

访问文档库

我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.

联系技术服务部门