跳转至内容
Merck
所有图片(1)

主要文件

G003766

HPLC Analysis of Diazepam and Metabolites in Serum on Discovery® C18 after SPE using Discovery® DSC-8

application for SPE, application for HPLC

登录查看公司和协议定价


About This Item

技术

HPLC: suitable
solid phase extraction (SPE): suitable

试验参数

sample preparation: SPE (Solid Phase Extraction)
sample/matrix: goat serum spiked with diazepam and metabolites at 0.5 μg/mL and diluted 1:1 in 25 mM ammonium formate, pH 7.1
SPE well plate: Discovery DSC-8 96-well SPE Plate, 100 mg/well (575627-U)
condition: 1 mL methanol followed by 1 mL DI water.
sample addition: 1 mL spiked goat serum
washing: 1 mL 25% methanol in 25 mM ammonium formate, pH 2.75
elution: 1 mL 60% methanol in 25 mM ammonium formate, pH 2.75
elution: Evaporate eluate with nitrogen purge (30 °C; ~15 min.). Reconstitute in 200 μL mobile phase.
column: Discovery C18, 5 cm x 4.6 mm I.D., 5 μm particles (504947)
mobile phase: [A] Methanol; [B]10 mM ammonium acetate, pH 4.5;(45:55, A:B)
flow rate: 1.5 mL/min
column temp.: 35 °C
detector: UV, 240 nm
injection: 25 μL

适用性

application for HPLC
application for SPE

应用

clinical
forensics and toxicology

法律信息

Discovery is a registered trademark of Merck KGaA, Darmstadt, Germany

历史批次信息供参考:

分析证书(COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our 文件 section.

如需帮助,请联系 客户支持

已有该产品?

在文件库中查找您最近购买产品的文档。

访问文档库

我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.

联系技术服务部门