Reagents

Prod. No. 104003 Formaldehyde solution min. 37 % for analysis
Prod. No. 109099 Iodine solution 0.05 mol I2/l 0.1 N solution
Prod. No. 109147 Sodium thiosulfate solution 0.1 mol/l 0.1 N solution
Prod. No. 109137 Sodium hydroxide solution 1 mol/l 1 N solution
Prod. No. 100716 Sulfuric acid 25 % for analysis
Prod. No. 105445 Zinc iodide-starch solution for analysis
Prod. No. 116754 Water for analysis

Preparation of a Stock Solution

In a calibrated or conformity-checked 1000-mL volumetric flask make up 2.50 mL of formaldehyde solution min.
37 % to the mark with distilled water. The stock solution prepared according to this procedure has a concentration of approximately 1000 mg/l formaldehyde.

Precise Assay of the Standard Solution (Stock Solution) for Formaldehyde

Pipette 40.0 mL (full pipette) of the formaldehyde stock solution of approximately 1000 mg/l into a 300-mL ground-glass conical flask, and add 50.0 mL (buret) of iodine solution 0.05 mol/l and 20 mL of sodium hydroxide solution 1 mol/l. Leave to stand for 15 minutes and subsequently add 8 mL of sulfuric acid 25 %. Add 1 mL of zinc iodide-starch solution and subsequently titrate to the end point with sodium thiosulfate solution 0.1 mol/l.

Calculation of the Exact Content of the Formaldehyde Solution

C1 = consumption of sodium thiosulfate solution 0.1 mol/l
C2 = quantity of iodine solution 0.05 mol/l
mg/l Formaldehyde = (C2 - C1) x 37.525

Further investigational concentrations may be prepared from the stock solution exactly determined according to the procedure described above by diluting accordingly.

Stability