Skip to Content
MilliporeSigma

HPLC Analysis of Warfarin Enantiomers on Astec® CHIROBIOTIC® V (Method 1)

HPLC Analysis of Warfarin Enantiomers on Astec® CHIROBIOTIC® V (Method 1) application for HPLC

Materials

analytical column

Product No.
Description
Pricing

Astec® CHIROBIOTIC® V Chiral HPLC Column

5 μm particle size, L × I.D. 25 cm × 4.6 mm

Application

Product No.
Description
Pricing

HPLC Analysis of Warfarin Enantiomers on Astec® CHIROBIOTIC® V2

application for HPLC

HPLC Analysis of Warfarin Enantiomers on Astec® CHIROBIOTIC® V (Method 2)

application for HPLC

HPLC Analysis of Warfarin Enantiomers on Astec® Cellulose DMP

application for HPLC

HPLC Analysis of Warfarin Enantiomers on Astec® CYCLOBOND I 2000 (UV at 278 nm)

application for HPLC

standard

Product No.
Description
Pricing

Warfarin

analytical standard

Warfarin

PESTANAL®, analytical standard

(R)-(+)-Warfarin

≥97% (HPLC)

(S)-(−)-Warfarin

≥97% (HPLC)

CONDITIONS

column

CHIROBIOTIC V, 25 cm x 4.6 mm I.D., 5 μm particles (11024AST)

mobile phase

[A] acetonitrile; [B] 5 mM ammonium acetate, pH 4.1; (30:70, A:B)

flow rate

1 mL/min

column temp.

25 °C

detector

UV, 254 nm

injection

5 μL

sample

1 mg/mL in mobile phase

Description

Legal Information

Astec is a registered trademark of Merck KGaA, Darmstadt, Germany
CHIROBIOTIC is a registered trademark of Sigma-Aldrich Co. LLC