Skip to Content
MilliporeSigma

GC Analysis of Conjugated Linoleic Acid (CLA) FAME Isomers on SLB®-IL111

GC Analysis of Conjugated Linoleic Acid (CLA) FAME Isomers on SLB®-IL111 suitable for GC

Materials

Analytical column

Product No.
Description
Pricing

SLB®-IL111 Capillary GC Column

L × I.D. 200 m × 0.25 mm, df 0.20 μm

CONDITIONS

column

SLB-IL111, 200 m x 0.25 mm I.D., 0.20 μm (29689-U)

oven

150 °C

inj. temp.

250 °C

detector

FID, 250 °C

carrier gas

hydrogen

injection

1 μL 1μL, split 50:1

liner

4 mm I.D., split type, cup design

sample

4-component CLA FAME isomer mix, 0.5 mg/mL in methylene chloride

Description

Analysis Note

Four key conjugated linoleic acid (CLA) FAME isomers are resolved. The SLB-IL111 provides a different elution pattern compared to SP-2560 (a cyanopropyl siloxane column).

Legal Information

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany