Tuerk's solution is not suitable for counting thrombocytes/platelets. It is advisable to use phase contrast microscopy for platelet counting, as Tuerk's solution may destroy platelets.
Wybierz wielkość
| Rozmiar/SKU | Dostępność | Cena netto |
|---|---|---|
100 mL | Skontaktuj się z Obsługą Klienta, aby uzyskać informacje na temat dostępności | 111,00 zł |
Informacje o tej pozycji
111,00 zł
Quality Segment
form
liquid
IVD
for in vitro diagnostic use
technique(s)
microbe id | staining: suitable
pH
2.5-3.5
density
1.00 g/cm3 at 20 °C
application(s)
clinical testing
diagnostic assay manufacturing
hematology
histology
storage temp.
15-25°C
General description
The basis of all counting methods is the dilution and preparation of a blood sample of known volume. The erythrocytes are hemolyzed by the acetic acid present in Tuerk′s solution and the leukocyte nuclei are stained blue-violet by the dye. The required cell type in a defined volume is counted and the number of cells per microliter of blood is then calculated.
The 100ml bottle of Tuerks solution will provide approx. 10 applications. Tuerk′s solution is an IVD product and CE registered.For more details, please see instructions for use (IFU).
Klasa składowania
12 - Non Combustible Liquids
wgk
WGK 1
Wykazy regulacyjne
Wykazy regulacyjne dotyczą głównie produktów chemicznych. Można w nich podawać ograniczoną liczbę informacji na temat produktów niechemicznych. Brak wpisu oznacza, że żaden ze składników nie znajduje się w wykazie. Użytkownik odpowiada za zagwarantowanie bezpiecznego i zgodnego z prawem stosowania produktu.
64-19-7
Wybierz jedną z najnowszych wersji:
Masz już ten produkt?
Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.
-
Can Tuerk's solution stain platelets/thrombocytes? Is there any other better stain for counting platelets?
1 answer-
Helpful?
-
-
what is the concentration of Turks solution
1 answer-
The concentration of the Turks solution is not determined.
Helpful?
-
-
Can I count PBMC with Turk Solution? Or what are the best option for PBMC counting? and another question Can Turk stain Thrombocytes?
1 answer-
Tuerk's Solution is primarily used for counting the total white cells present in a peripheral blood specimen, including PBMCs (lymphocytes and monocytes).
For accurate counting of PBMCs, the most common approach is to use flow cytometry or hemocytometers. Hematology labs will typically have dedicated cell counters that will also count white blood cells and also classify the percentage of each type of white blood cell found in the blood specimen. Alternatively, instruments, such as the Scepter™ cell counter, will differentiate PBMCs based on size.See the link below to review an application note that may be of interest:
https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/product/documents/326/775/an3311en-ms.pdfTuerk's solution is not routinely used for counting thrombocytes or platelets. Although some platelets may show some staining, there will be platelets that remain unstained. Most labs that count platelets manually will use ammonium oxalate to lyse the red blood cells and then use a phase hemacytometer and a phase (contrast) microscope.