HPLC Analysis of Fat Soluble Vitamins by Normal Phase Chromatography on Ascentis® Si (25 cm Column)

application for HPLC

Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización


About This Item

technique(s)

HPLC: suitable

test parameters

column: Ascentis Si, 15 cm x 4.6 mm I.D., 5 μm particles (581512-U)
mobile phase: [A] hexane; [B] ethyl acetate
gradient: 10 to 30% B in 10 min; held at 30% B for 2 min
flow rate: 1.0 mL/min
column temp.: 30 °C
detector: UV, 290 nm
injection: 10 μL

suitability

application for HPLC

application(s)

vitamins, nutraceuticals, and natural products

Analysis Note

Ascentis silica is used in the analysis of fat-soluble vitamins using a normal phase gradient. These include vitamin K 3 (menadione), vitamin E (alpha and gamma tocopherol), and vitamin D3 (cholecalciferol).

Legal Information

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analito

Descripción

    (±)-α-Tocopherol, analytical standard
    Menadione (K3), analytical standard
    (+)-γ-Tocopherol, analytical standard
    Cholecalciferol (D3), analytical standard

analytical column

Referencia del producto
Descripción
Precios

standard

Referencia del producto
Descripción
Precios

Elija entre una de las versiones más recientes:

Certificados de análisis (COA)

Lot/Batch Number

Lo sentimos, en este momento no disponemos de COAs para este producto en línea.

Si necesita más asistencia, póngase en contacto con Atención al cliente

¿Ya tiene este producto?

Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.

Visite la Librería de documentos

Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.

Póngase en contacto con el Servicio técnico