HPLC Analysis of a Monoclonal Antibody on Zenix® SEC-150, 30 cm X 4.6 mm I.D., 3 μm particles after Proteolysis with IdeS Protease
application for HPLC
About This Item
technique(s)
HPLC: suitable
test parameters
column: Zenix® SEC-150, 30 cm X 4.6 mm I.D., 3 μm particles (Z777018)
mobile phase: 0.15 M potassium phosphate monobasic, pH 7.0 with potassium hydroxide
flow rate: 0.35 mL/min
pressure: 1450 psi (100 bar)
column temp.: ambient
detector: UV, 215 nm
injection: 1 μL
sample: digested SO57 mAb, ~0.5 g/L
suitability
application for HPLC
application(s)
life science and biopharma
Analysis Note