Saltar al contenido
Merck
Todas las fotos(1)

Documentos

G005512

EN 14110: GC Analysis of Methanol Impurity in Biodiesel on Equity-1 after SPME using 85 μm Polyacrylate Fiber

suitable for GC, application for SPME

Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización


About This Item

technique(s)

gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable

test parameters

sample/matrix: 2 g of B100 biodiesel containing 0.01% methanol and 0.0785% 2-propanol in a 4 mL vial
SPME fiber: PK3 SPME ASSY 85UM POLYACRYLATE FS 24GA MAN (WHITE) (57304)
extraction: headspace, 60 °C for 5 min
desorption process: 250 °C for 0.75 min
column: Equity-1, 30 m x 0.32 mm I.D., 1.0 μm (28057-U)
oven: 50 °C
inj. temp.: 250 °C
detector: FID, 200 °C
carrier gas: helium, 30 cm/sec
liner: 0.75 mm I.D., SPME type, straight design (unpacked)

suitability

application for SPME

application(s)

biofuels
environmental

Legal Information

SPME fiber

Referencia del producto
Descripción
Precios

analytical column

Referencia del producto
Descripción
Precios

Elija entre una de las versiones más recientes:

Certificados de análisis (COA)

Lot/Batch Number

Lo sentimos, en este momento no disponemos de COAs para este producto en línea.

Si necesita más asistencia, póngase en contacto con Atención al cliente

¿Ya tiene este producto?

Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.

Visite la Librería de documentos

Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.

Póngase en contacto con el Servicio técnico