G005293
GC Analysis of Trichloroanisole (TCA) and Precursors on SLB®-5ms after SPME using 100 μm a PDMS Fiber
suitable for GC, application for SPME
About This Item
technique(s)
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
test parameters
sample/matrix: each analyte at 100 ng/L in a 12% ethanol solution
SPME fiber: metal fiber coated with 100 μm PDMS (57928-U)
extraction: headspace, 50 °C for 30 min, with stirring
desorption process: 250 °C for 3 min
column: SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (28471-U)
oven: 50 °C (1 min), 25 °C/min to 280 °C
inj. temp.: 250 °C
detector: GC/MS, ECD, 290 °C
carrier gas: helium, 1.5 mL/min, constant
injection: splitless/split, closed 2 min
liner: 0.75 mm I.D. SPME
sample: Wine cork
suitability
application for SPME
application(s)
food and beverages
Analysis Note
Legal Information
Analito
SPME fiber
analytical column
standard