The protocol for cell samples using MAK434 is as follows:
- Suspend about two million (2 × 10^6) harvested cells in 400 μL PBS on ice. Lysis can be achieved by homogenization (10-20 passes in a Dounce homogenizer on ice) or by sonication (preferably performed in an ice-water bath). The degree of cell lysis can be checked under a microscope. Centrifuge the homogenate at 14,000 g for 10 min and transfer the clear supernatant into a clean tube. It is advisable to run a pilot test of the sample at different dilutions and select a dilution with readings in the linear range of the standard curve for further assays. Most samples can be stored at -80°C if not assayed immediately.
- If a lysis buffer is to be used, it is recommended to first run an assay to test if components in the lysis buffer interfere with the assay. This can be accomplished by running two standard curves: one prepared in H2O and one prepared in the same proportion of lysis buffer that the sample will contain. If the two standard curves are the same, there is no interference and it is safe to use the lysis buffer. If there is a difference in the standard curves, the lysis buffer may still be used if the standard curve remains linear; however, the standard curve to be used for analyzing the lysate samples will need to be prepared in the lysis buffer.
MAK434
Acetaldehyde Assay Kit
Sufficient for 100 colorimetric tests
Seleccione un Tamaño
Seleccione un Tamaño
About This Item
Productos recomendados
Nivel de calidad
entrada
cell culture medium
tissue
serum
urine
food(s)
beverage(s)
plasma
aplicaciones
agriculture
cosmetics
environmental
food and beverages
método de detección
colorimetric
enfermedades relevantes
gastrointestinal diseases
temp. de almacenamiento
−20°C
Descripción general
Aplicación
- Alcohol Metabolism Research
- Hepatology Research
- Neurological Research
Características y beneficios
- Simplified Process: Enjoy a hassle-free procedure that requires the addition of a single working reagent and only a 30 minute assay room temperature reaction, saving you time and effort.
- Convenient and Sensitive: Uses only 20 µL of sample for the determination of acetaldehyde with a detection range of 2 µM to 2mM.
- High-Throughput Compatibility: Seamlessly integrate our kit into high-throughput handling systems, ensuring efficiency and accuracy.
Idoneidad
Principio
Otras notas
Palabra de señalización
Warning
Frases de peligro
Consejos de prudencia
Clasificaciones de peligro
Eye Irrit. 2 - Skin Irrit. 2
Código de clase de almacenamiento
12 - Non Combustible Liquids
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
¿No ve la versión correcta?
Si necesita una versión concreta, puede buscar un certificado específico por el número de lote.
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
-
How should MAK434 be used for cell samples?
1 answer-
Helpful?
-
Active Filters
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico