64012BC
MVP ICON® ATP Calibrator Set
MVP ICON®, ATP Luminometer Calibration
Synonym(s):
ATP Calibrator Set, ATP Measurement Calibration Kit, Quality Control ATP Calibrator
Sign Into View Organizational & Contract Pricing
All Photos(2)
About This Item
Recommended Products
Agency
EP
ISO
USP
product line
MVP ICON®
shelf life
5 yr
feature
ready-to-use
manufacturer/tradename
MVP ICON®
packaging
set of calibrator rods
storage condition
protect from light (Keep in a tightlyclosed container, away from any light sources)
application(s)
food and beverages
compatibility
for use with MVP ICON instrument
storage temp.
15-25°C
General description
MVP ICON® ATP Calibrator Set consists of two calibration devices (“rods”) that are used for performance verification and calibration of the MVP ICON® and MVP systems. The luminescent calibrator consists of a 14C radioactive source that emits a very low level (5 x 105 d.p.m., or 0.225 microcuries) of low-energy ß-radiation in a plastic scintillation matrix. This matrix is configured to give constant light output, within 10 % of its original value, at a sufficiently high level to make consistent verification and calibration functions possible over the five‐year life of the calibrator. The non‐luminescent black plastic calibrator provides a reliable low‐end or “dark” source reading.
Application
MVP ICON® ATP Calibrator Set is used for performance verification and calibration of the MVP ICON® instrument for rapid monitoring of surface hygiene in the Food and Beverage industries.
Features and Benefits
- On-site calibration
- Satisfies HACCP and GMP requirements
- No need to send the equipment out to a 3rd party.
Components
- One luminescent calibrator
- One non-luminescent calibrator
Suitability
ATP Testing, Food Processing, HACCP Monitoring, Sanitation
Principle
The luminescent calibrator consists of a 14C radioactive source that emits a very low level (5 x 105 d.p.m., or 0.225 microcuries) of low energy ß radiation in a plastic scintillation matrix. This matrix is configured to give constant light output, within 10% of its original value, at a sufficiently high level to make consistent verification and calibration functions possible over the five- year life of the calibrator. The non-luminescent black plastic calibrator provides a reliable low-end or “dark” source reading.
Testing Method
ATP Bioluminescence
Storage and Stability
To ensure the stability and longevity of the luminescent calibrator, the following precautions must be taken:
- Limit Exposure to Light.
- Store the luminescent calibrator in the dark or weak ambient light.
- If the luminescent calibrator is exposed to bright light before use, insert it in the MVP ICON® or MVP and wait a minimum of 2 minutes before proceeding.
- The kit box provided is recommended as an appropriate dark storage.
- Do not expose to direct sunlight or intense artificial light before use.
Other Notes
Avoid crushing, cutting, or dropping the calibrators.
Legal Information
BioControl is a registered trademark of BioControl Systems, Inc.
MVP ICON is a registered trademark of Merck KGaA, Darmstadt, Germany
Kit Components Also Available Separately
Product No.
Description
SDS
- 72001SBCLuminescent Calibrator RodSDS
- Non Luminescent Calibrator Rod
Storage Class Code
10 - Combustible liquids
Flash Point(F)
Not applicable
Flash Point(C)
Not applicable
Choose from one of the most recent versions:
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Customers Also Viewed
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service