G005922
HPLC Analysis of Aflatoxins B1, B2, G1 and G2 in Corn on Ascentis® Express C18 after SPE using Supel™ Tox AflaZea and Fluorescence Detection
application for HPLC
About This Item
Technique(s)
HPLC: suitable
Paramètres du test
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: 25 g corn meal in 100 mL acetonitrile:water (84:16)
SPE tube/cartridge: Supel™ Tox AflaZea, 6 mL (55314-U)
sample addition: 2 mL
elution: collect sample in 5 mL silanized glass vial
eluate post-treatment: transfer 200 μL sample, add 300 μL TFA derivatization solution (70:20:10 water:TFA:acetic acid), heat in microreactor 20 min at 65 °C, cool, add 580 μL deionized water, vortex
column: Ascentis Express C18, 15 cm x 2.1 mm I.D., 2.7 μm particles (53825-U)
mobile phase: [A] water; [B] acetonitrile; (80:20, A:B)
flow rate: 0.350 mL/min
pressure: 325 psi (22.4 bar)
column temp.: 35 °C
detector: fluorescence, ex 360 nm, em 440 nm
injection: 20 μL
Adéquation
application for HPLC
Application(s)
food and beverages
Remarque sur l'analyse
Informations légales
Analyte
Colonne analytique
Produit(s) apparenté(s)
Faites votre choix parmi les versions les plus récentes :
Certificats d'analyse (COA)
Désolés, nous n'avons pas de COA pour ce produit disponible en ligne pour le moment.
Si vous avez besoin d'assistance, veuillez contacter Service Clients
Déjà en possession de ce produit ?
Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.
Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..
Contacter notre Service technique