G005284
GC Analysis of Glycols in Water on SPB®-1000 after SPE using ENVI-Carb™ Plus
suitable for GC, application for SPE
About This Item
Empfohlene Produkte
Methode(n)
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
Testparameter
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: tap water spiked at 2 μg/mL with glycols
SPE tube/cartridge: ENVI-Carb Plus reversible cartridge, 400 mg/ 1mL (54812-U)
condition: 1 mL methylene chloride, 2 x 2 mL methanol, 3 mL deionized water (cartridge not allowed to dry during conditioning)
sample addition: 5 mL spiked tap water
drying: 10 min, 10 mm mercury
elution: 2 mL 80:20 methanol:methylene chloride
column: SPB-1000, 30 m x 0.25 mm I.D., 0.25 μm (24313)
oven: 50 °C (1 min), 8 °C/min to 200 °C (7 min)
inj. temp.: 220 °C
detector: FID, 220 °C
carrier gas: helium, 1.5 mL/min. constant
injection: 1 μL, splitless (splitter open at 0.75 min.)
liner: 4 mm I.D. FocusLiner™ inlet liner with taper
Eignung
application for SPE
Anwendung(en)
environmental
Rechtliche Hinweise
Analyt
Analytische Säule
SPE-Röhrchen oder -Platte
Hier finden Sie alle aktuellen Versionen:
Analysenzertifikate (COA)
Leider sind derzeit keine COAs für dieses Produkt online verfügbar.
Wenn Sie Hilfe benötigen, wenden Sie sich bitte an Kundensupport
Besitzen Sie dieses Produkt bereits?
In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.
Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..
Setzen Sie sich mit dem technischen Dienst in Verbindung.