콘텐츠로 건너뛰기
Merck

UHPLC Analysis of Fat Soluble Vitamins on Titan C18, Acetonitrile Mobile Phase

UHPLC Analysis of Fat Soluble Vitamins on Titan C18, Acetonitrile Mobile Phase application for UHPLC

Materials

표준 물질

제품 번호
설명
가격

DL-α-Tocopherol acetate

≥96% (HPLC)

Retinyl acetate

analytical standard

(±)-α-Tocopherol

analytical standard

(+)-γ-Tocopherol

analytical standard

δ-Tocopherol

analytical standard

CONDITIONS

column

TITAN C18, 50 mm x 2.1mm I.D. x 1.9 μm particles (577122-U)

mobile phase

100% acetonitrile

flow rate

0.4 mL/min

pressure

1755 psi (121 bar)

column temp.

30 °C

detector

UV, 280 nm

injection

0.5 μL

sample

All compounds at 0.5 mg/mL except retinyl acetate at 0.05 mg/mL in methanol