콘텐츠로 건너뛰기
Merck

EN 14110: GC Analysis of Methanol Impurity in Biodiesel on Equity-1 after SPME using 85 μm Polyacrylate Fiber

EN 14110: GC Analysis of Methanol Impurity in Biodiesel on Equity-1 after SPME using 85 μm Polyacrylate Fiber suitable for GC, application for SPME

Materials

SPME 파이버

제품 번호
설명
가격

SPME fiber assembly polyacrylate (PA)

df 85 μm(PA, for use with manual holder, needle size 24 ga

분석 컬럼

제품 번호
설명
가격

Equity-1 Capillary GC Column

L × I.D. 30 m × 0.32 mm, df 1.00 μm

CONDITIONS

sample/matrix

2 g of B100 biodiesel containing 0.01% methanol and 0.0785% 2-propanol in a 4 mL vial

SPME fiber

PK3 SPME ASSY 85UM POLYACRYLATE FS 24GA MAN (WHITE) (57304)

extraction

headspace, 60 °C for 5 min

desorption process

250 °C for 0.75 min

column

Equity-1, 30 m x 0.32 mm I.D., 1.0 μm (28057-U)

oven

50 °C

inj. temp.

250 °C

detector

FID, 200 °C

carrier gas

helium, 30 cm/sec

liner

0.75 mm I.D., SPME type, straight design (unpacked)